To calibrate infrared pyrometers and thermal imaging systems accurately, it is necessary to have a source of IR radiance that is controlled and predictable. PSC-BBS series of blackbody calibration sources provide a target of known emissivity, temperature and aperture size. These blackbodies can also provide calibration traceability to certified laboratories and national standards.
Download the Blackbody Source Overview brochure
Blackbody Product Series / Mid to High Range
PDF |
PRODUCT |
TEMP. RANGE |
APERTURE SIZE |
EMISSIVITY VALUE |
POWER REQUIREMENT |
|
PSC-BBS-1050 Compact
|
50° to 1050°C or 122° to 1922°F |
1" max, adjustable to 7 other smaller sizes |
>0.99 |
90 to 125VAC or 208 to 240VAC (must select one)
50/60Hz, 550 W max |
|
PSC-BBS-1200 Compact
|
50° to 1200°C or 122° to 2192°F) |
1" max, adjustable to 7 other smaller sizes |
>0.99 |
90 to 125VAC or 208 to 240VAC (must select one)
50/60Hz, 800 W max |
|
PSC-BBS-1150
|
200° to 1150°C or
392° to 2102°F |
2" (51mm) |
+0.995 |
115VAC/230VAC 60Hz
2kw max. |
|
PSC-BBS-1500
|
500° to 1500°C |
1” (25mm) |
+0.99 |
220VAC, 50/60Hz, 3.8 KW (21 amps) |
|
PSC-BBS-1700
|
300° to 1700°C or
572° to 3902°F |
1" (25mm) |
+0.99 |
208VAC ±10% 50/60Hz
3.0kw max. (230VAC optional) |
|