Relative humidity measurement
|
|
Measurement range |
0 ... 100 % RH |
|
Accuracy at +20 °C (68 °F) |
± 1.0 % RH (0 ... 90 % RH)
± 1.7 % RH (90 ... 100 % RH) |
|
Sensor |
Vaisala HUMICAP® 180 |
|
general use |
Vaisala HUMICAP® 180R |
|
applications with a
demanding chemical
environment |
Vaisala HUMICAP® 180L2 |
Temperature measurement
|
|
Measurement range
|
|
|
HMT361 |
-40 ... +60 °C (-40 ... +140 °F) |
|
HMT363 |
-40 ... +120 °C (-40 ... +248 °F) |
|
HMT364/365/367/368 |
-40 ... +180 °C (-40 ... +356 °F) |
|
Typical accuracy of
electronics at +20 °C
(68 °F) |
± 0.2 °C (± 0.36 °F) |
|
Sensor |
Pt 1000 RTD 1/3 Class B IEC 751 |
Other optional variables available
|
|
Dewpoint temperature, mixing ratio, absolute humidity, wet bulb temperature.
|
Operating environment
|
|
Temperature
for electronics
with display
storage
|
-40 ...+ 60 °C (-40 ... +140 °F)
-20 ...+ 60 °C (-4 ... +140 °F)
-40 ...+ 70 °C (-40 ... +158 °F)
|
Inputs and outputs
|
|
Operating voltage |
12 ... 28 V |
|
with serial port (service
mode) |
15 ... 28 V |
|
Analog outputs |
two-wire 4 ... 20 mA, one standard, one optional |
|
Digital outputs |
RS-232C serial output for service use |
Mechanics
|
|
Weight (housing)
Material (housing)
Housing classification
|
950 g
G-AlSi 10 Mg (DIN 1725)
IP66 (NEMA 4X)
|
Options and accessories
|
|
Additional analog output |
4 ... 20 mA |
|
Duct mounting installation kit
(for HMP363/367) |
210697 |
|
Installation flange
(for HMP365) |
210696 |
|
Ball valve set (for HMP368) |
BALLVALVE-1 |
|
Serial interface cable for PC |
25905ZZ |
|
Protection cover |
214101 (for use in the presence of combustible dust) |
|
Galvanic isolator |
212483 |
|
Zener barrier (USA & Canada) |
210664 |