Relative humidity
|
|
Measurement range |
0 ... 100 %RH |
|
Accuracy including non-linearity, hysteresis, and repeatability
at 0 °C ... +40 °C |
±1.7 %RH (0 ... 90 %RH) |
|
|
±2.5 %RH (90 ... 100 %RH) |
|
at -40 °C ... +0 °C and
at +40 °C ... +80 °C
Factory calibration uncertainty at +20 °C (+68 °F) |
±3.0 %RH (0 ... 90 %RH)
±4.0 %RH (90 ... 100 %RH)
±1.5 %RH |
|
Humidity sensor |
Vaisala HUMICAP® 180R |
Temperature
|
|
Measurement range |
-40 ... +80 °C |
|
Accuracy over temperature range
at +15 °C ... +25 °C
at 0 ... +15 °C and at +25 °C ... +40°C
at -40 °C ... +0 °C and at +40 °C ... +80 °C |
±0.2 °C
±0.25 °C
±0.4 °C |
|
Temperature Sensor
Operating temperature range
transmitter body, no display
transmitter body, with display
HMP110 probe
Storage temperature range
Electromagnetic compatibility |
Pt1000 RTD 1/3 Class B IEC 751
-40 °C ... +60 °C
-20 °C ... +60 °C
-40 °C ... +80 °C
-50 °C ... +70 °C
EN 61326-1
and EN 55022 |
Inputs and outputs
HMT120 TWO-WIRE TRANSMITTER (LOOP POWERED) |
|
Current output signals
External loop voltage
HMT130 THREE-WIRE TRANSMITTER
Voltage output signals
Min output resistance
Serial output
Relay output
Supply voltage
Current consumption at 24 VDC
Max. additional error caused by the |
4 ... 20 mA
10 ... 30 VDC (RL= 0 ohms)
20 ... 30 VDC (RL< 500 ohms)
0 ... 1 V, 0 ... 5 V, 0 ... 10 V or
user defined between 0 ... 10 V
1 kohm
RS485, non-isolated
1 relay (max. 50 VDC, 200 mA)
10 ... 35 VDC
15 ... 35 VDC (when output 0 ... 10 V)
24 VAC (±20 %)
8 mA, if relay closed 15 mA
±0.1 % of FS output signal |
analog outputs after calibration at
+20 °C ambient temperature |
|
Temperature dependence of the |
±0.005 % of FS output signal |
analog outputs |
|
Mechanics
Materials |
|
|
transmitter housing
display window |
PBT plastic
PC plastic |
|
probe body |
Chrome coated aluminum |
|
probe grid filter |
Chrome coated ABS plastic |
|
Housing classification |
IP65 |
|
Connections |
|
|
inputs and outputs |
Screw terminals 0.5 ... 1.5 mm2 |
|
probe interface
Probe cable length
Display (optional)
Weight (including probe) |
4-pin M8 female panel connector
3 m, 5 m, 10 m - up to 50 m
128 x 64 resolution full graphics
B&W display without backlight
270 g |
Accessories
|
|
Humidity and temperature probe
Humidity and temperature replacement probe
Constant output probe
Humidity sensor
Probe mounting flange
Probe mounting clamps, 10 pcs |
HMP110*
HMP110R*
HMP110REF*
HUMICAP® 180R
226061
226067 |
HMP110 sensor protection |
|
plastic grid filter |
DRW010522SP |
|
plastic grid with membrane filter
stainless steel sintered filter |
DRW010525
HM46670SP |
|
Probe cable 3 m |
HMT120Z300 |
|
Probe cable 5 m |
HMT120Z500 |
|
Probe cable 10 m
Radiation shield
Probe installation kit for DTR502
Rain shield with installation kit
Duct installation kit
HMI41 connection cable
HM70 connection cable
USB serial interface cable |
HMT120Z1000
DTR502B
210623
215109
215619
25917ZZ
211339
219685 |