Vaisala HUMICAP Moisture and Temperature Transmitter for Oil MMT162
|
- Compact, accurate and economical
- Proven Vaisala HUMICAP® technology
- Continuous measurement of moisture in oil
- Excellent pressure and temperature tolerance
- Compatible with the Vaisala MI70 indicator
|
Measured Values
|
|
Water activity
|
|
|
Measurement range |
0 ... 1 aw |
|
Accuracy (including non-linearity, hysteresis and repeatability) |
|
|
0 ... 0.9 |
± 0.02 |
|
0.9 ... 1.0 |
± 0.03 |
|
Response time
in oil flow (typical) |
<1 min (dry-wet) |
|
|
|
|
Moisture |
Calculated moisture content in ppm for mineral transformer oil |
|
Temperature |
|
|
Accuracy at +20 °C (+68 °F) |
± 0.2 °C (± 0.36 °F) |
Operating Environment
|
|
Operating temperature |
-40 ... +60 °C (-40 ... +140 °F) |
|
Oil temperature |
-40 ... +80 °C (-40 ... +176 °F) |
|
Pressure range |
|
|
metal version |
up to 200 bar |
|
plastic version |
up to 40 bar |
|
Oil flow |
some flow recommended |
Outputs
|
|
Analog outputs (two channels) |
|
|
current output |
0 ... 20 mA, 4 ... 20 mA |
|
voltage output |
0 ... 5 V, 0 ... 10 V |
|
Digital outputs |
RS-485 |
General
|
|
Housing material |
|
|
metal |
AISI 316L |
|
plastic |
PPS + 40% GF |
|
Mechanical connections with bonded seal ring (washer) |
|
|
metal version |
G 1/2’’ ISO or NPT 1/2’’ |
|
plastic version |
G 1/2’’ ISO |
|
|